Figure 1 Model 32301 sensor cleaning and regeneration center
1 120 or 230Vac power input
2 Circuit test sockets
3 Sensor cable for detachable sensors
4 Selector knob
The unit has a timer switch (No. 5) which switches off automatically after 60 seconds, a selector knob
(No. 4) and a counter electrode socket (No. 7). The selector knob offers the choice of cleaning or
chloridizing functions. For oxygen or ozone sensors, the anode, cathode and guard ring electrode
cleaning positions are available. The two selector positions in the chloridizing section cover
dechloridization and rechloridization of hydrogen sensors. The counter electrode socket takes the
banana plug from the regeneration cell.
For electrical testing there are eleven sockets (No. 2) to accept banana plugs. Each of the sockets is
connected to one of the ten pins of a sensor LEMO plug (including a shield plug). By connecting an
ohmmeter across two of the sockets sensor resistances can be measured and the state of the sensor
checked.
Figure 2 Regeneration cell for 31xxx series
Use the supplied regeneration cell 29037 to clean 31xxx series. Refer to
banana plug into the counter electrode socket on the sensor cleaning and regeneration center.
6 English
5 Timer switch
6 Plug for sensors with attached cable
7 Connects to regeneration cell