GF U3000 V2 Betriebsanleitung Seite 61

Ultraschall-durchflussmessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 73
User Manual
Battery errors
Battery critically low!
BATTERY EXHAUSTED! Shut down in
[n] sec!
Set-up and other errors & messages
Too many errors
Too many urgent errors
Poorly formatted error message
Site DB is full
Site name illegal or duplicate
Energy calculations unreliable
RTD Board fault
Power Board fault
Logger Board fault
Output Board fault
Flow Board fault
Limits are xx.x [text] to yy.y [text]
Site DB failure. Restoring default values.
Code is invalid
Interpretation: URGENT: The internal battery voltage is less than 6.1
volts.
Response: Connect the external charger. Press 'delete' to remove
this error.
Interpretation: The internal battery voltage is less than 5.25 volts.
The flowmeter will perform a controlled shutdown in 15 seconds
if an external charger is not connected. The time to shutdown is [n]
seconds.
Response: Connect the external charger. Press 'delete' to remove
this error.
Interpretation: The flowmeter generated too many errors as a result
of a fault and some errors may not have been reported.
Response: Respond to the errors highlighted.
Interpretation: The flowmeter has generated too many urgent errors
as a result of a fault and some errors may not have been reported.
Response: Delete urgent errors before continuing. Urgent errors are
displayed before normal errors, so they are cleared first by pressing
the 'delete' key.
Interpretation: Internal, NON- FATAL System error. Response: Delete
the error. Take note of the current situation that lead to this error and
report it when convenient.
Interpretation: The number of sites has exceeded the maximum of
12.
Response: Delete a site.
Interpretation: Site names must be unique and contain eight or fewer
characters comprising letters, numbers dashes or underscores.
Response: Enter a site name that complies with the interpretation
above. Note that names are case insensitive, for example, site ELY is
a duplicate of site Ely.
Interpretation: The temperature used in heat-meters calculations is
outside that which can be accurately calculated.
Response: This is a NON-FATAL error. If the error is persistent, check
your installation for temperatures that are out-of-bounds, and check
the leads to the
temperature probes.
Interpretation: URGENT: The respective board has not reported to the
central controller in the last minute.
Response: Try restarting the flowmeter. If the board is still reported
as missing or faulty, call your distributor or return your device for
repair. You may press 'delete' to remove this error but some or all
functionality may be lost if this error is persistent and you continue
operate the device.
Interpretation: The values entered were out of bounds for this
setting. The smallest value allowed is xx.x and the largest is yy.y.
Optional units [text] may accompany this message. If not, then it is
implied that the units are those
currently set.
Response: Enter a value within the specified limits. Note that the
limits quoted may be dependent on other parameters already set.
Interpretation: When reading parameters from the database, some
site parameters appeared corrupted, so all parameters have been
restored to initial values. Response: Re-enter parameters for this
site. Press 'delete' to remove this error.
Interpretation: Either the user or factory pin code is incorrect.
Response: Try again.
Troubleshooting
61