Optional accessories
Technical Data
(RBW 100 Hz, VBW 10 Hz, ref.
level ≤-30 dBm 10 MHz to
1.6 GHz/3 GHz*1)
Preamp. deactivated
Inherent spurious:
(ref.-level ≤-20 dBm,
f >30 MHz, RBW ≤100 kHz)
Input related spurious:
)
(Mixer level ≤-40 dBm,
carrier offset >1 MHz)
(2 to 3 GHz)
2nd harmonic receive frequency:
(mixer level -40 dBm)
Level display:
Reference level:
Display range:
Logarithmic display scaling:
Linear display scaling:
Measured curves:
Trace mathematics:
Detectors:
Failure of level display:
Marker/Deltamarker
Number of marker:
Marker functions:
Marker displays:
Inputs/Outputs
HF Input:
Input impedance:
VSWR
(10 MHz to 1,6 GHz/3 GHz
Output tracking generator
Output impedance:
Frequency range:
Output level:
Trigger input:
Trigger voltage:
Ext. reference input/output:
Reference frequency:
Essential level (50 Ω):
Supply output for field probes:
Audio output (Phone):
Demodulation
Miscellaneous
Display:
Save/Recall memory:
Trigger:
1) with HMS-3G (HV212) option
2) with HMS-EMC (HV213) option
3) with HMS-TG (HV211) option
74
-115 dBm
2
), typ. -135 dBm
typ. -124 dBm
2
)
<-80 dBm
-70 dBc typ.
-55 dBc typ.
1
)
-60 dBc typ.
-80 bis +20 dBm in 1 dB steps
100 dB, 50 dB, 20 dB, 10 dB linear
dBm, dBµV, dBmV
Percentage of reference level
1 curve and 1 memory curve
A-B (curve-stored curve), B-A
Auto-, Min-, Max-Peak, Sample, RMS,
Average, Quasi-Peak
2
)
<1,5 dB, typ. 0,5 dB
(ref.-level -50 dBm, 20 to 30 °C)
8
Peak, next peak, minimum,
center = marker, frequency,
reference level = marker level,
all marker on peak
Normal (level, log.), delta marker,
noise marker, Normal (lin.), (frequency)
counter
2
)
N socket
50 Ω
1
)
<1,5 typ.
3
):
N socket
50 Ω
5 MHz to 1,6 GHz/3 GHz
-20 dBm to 0 dBm, in 1 dB steps
BNC female
TTL
BNC female
10 MHz
10 dBm
6 V
, max. 100 mA (2,5 mm DIN jack)
DC
3,5 mm DIN jack
AM and FM (internal speaker)
16,5 cm (6,5") TFT Color VGA Display
10 complete device settings
Free run, Single Trigger, external
Trigger, Video-Trigger
2
)
Interfaces:
2
)
Power supply:
Power consumption:
Protection class:
Operating temperature:
Storage temperature:
Rel. humidity:
Dimensions (W x H x D):
Weight:
)
2
2
)
Accessories included:
Line cord, printed operating manual, CD, software
Recommended accessories:
1
)
HO730 Dual-interface ethernet/USB
HO740 Interface IEEE-488 (GPIB), galvanically isolated
HZ530 Near-fi eld probe set 1 GHz for EMI diagnostics
HZ540/550
Near-fi eld probe set 3 GHz for EMI diagnostics
HZ540L/550L
Near-fi eld probe set 3 GHz for EMI diagnostics
HZ547 3 GHz VSWR bridge for HMS-X
incl. HMS-TG option
HZ13
Interface cable (USB) 1.8 m
HZ14
Interface cable (serial) 1:1
HZ21
Adapter N (plug) - BNC (socket)
HZ46
4RU 19" rackmount kit
HZ72
GPIB-cable 2 m
HZ99
Carrying case for protection and transport
HZ520 Plug-in antenna with BNC connection
HZ525 50 Ω-termination, N plug
HZ560 Transient limiter
HZ575 75/50 Ω converter
Dual-Interface USB/RS-232 (HO720),
USB-Stick (frontside),
USB-Printer (rear side),
DVI-D for ext. monitor
105/253 V, 50 to 60 Hz, CAT II
Max. 40 W at 230 V, 50 Hz
Safety class I (EN61010-1)
+5 to +40 °C
-20 to +70 °C
5 to 80 % (non condensing)
285 x 175 x 220 mm
3,6 kg