9 Menu
Access to service menu „x10" is locked in verified balances.
To disable the access lock, destroy the seal mark and actuate the
adjustment switch. For position of adjustment switch, see chap. 20.
Attention:
After destruction of the seal the weighing system must be re-verified by
an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.

9.1 Navigation in the menu

Call up menu
Select function
Change settings
Exit menu/
Return to weighing
mode
40
 In weighing mode press
„SetuP" followed by „coM" will appear.
 With help of
can be selected one after the other.
 Confirm selected function by
will be displayed.
 Select desired setting with
, the balance returns to the menu.
 Press
repeatedly until the zero display appears.
The balance is now in weighing mode.
simultaneously,
or
, the individual functions
. The current setting
and confirm by
TMPN_A-BAPS-e-2210