signals with different DC offsets (see also
on page 19).
To assign "Set offset to mean" to the micro button:
► "Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action" = "Set off-
set to mean"
The maximum input voltage is ±30 V. A higher input voltage may destroy
the probe.
2.4
R&S ProbeMeter
The integrated voltmeter can measure DC voltages with very high precision com-
pared to the oscilloscope's DC accuracy. The DC measurement is performed
continuously and in parallel to the time domain measurement of the oscilloscope.
High-precision measurements are achieved through immediate digitization of the
measured DC voltage inside the probe.
To activate the R&S ProbeMeter:
► Vertical > Probe Setup > Ch > R&S ProbeMeter 'on'
After the R&S ProbeMeter has been activated, the measured values are dis-
played on the screen of the oscilloscope.
Advantages of the R&S ProbeMeter:
● Enables the user to check DC voltages with different levels without having to
adjust the measurement range of the oscilloscope.
● Provides a simple means of setting the oscilloscope's trigger level and vertical
scaling if a waveform is not visible.
● Independent of oscilloscope settings for offset, position, vertical scale, hori-
zontal scale, and trigger.
● True DC measurement (integration time > 100 ms), not mathematical average
of displayed waveform.