4 CONFIGURATION
4.3.6 Relays
The relay thresholds are set in this menu option. For each relay, set 2 thresholds; 1- and 1+
respectively 2- and 2+. The operating direction of the relays can be inverted here. The unit
and number of decimal digits are those set within the "UNIT" menu option.
The following condition must be observed: 1- ≤ 1+, 2- ≤ 2+.
RELAY
1-=0000
ENTER
Setting low value of 1st
threshold.
INV NO
ENTER
INV YES
2-=0000
INV YES
ENTER
INV NO
FILTER
The digit can be incremented by successive presses on a key
0......9
or by pressing and holding a key.
1-=0008
1+=0000
ENTER
1+=0010
ENTER
0......9
2-=0040
2+=0000
ENTER
ENTER
F-36-
8025 / 8035 KONTI-DOS
Contact
invert no
0......9
closed
open
1- (8)
Contact
0......9
invert yes
2+=0044
closed
open
1- (8)
1+ (10)
Flow
1+ (10)
Flow