4 CONFIGURATION
4.3.3 K factor
This menu makes it possible to program the K factor of the connected sensor or S030
fitting (value between 0,01 and 80000). The "Teach in" procedure makes it possible to
experimentally determine the application specific K factor.This can be achieved by letting a
known amount of liquid flow through the installation.
Example: The user has a 100 litre capacity tank. When "TEACH Y" is displayed, press the
ENTER key to start measurement and switch on the pump (or open the valve). Message
"FILL END" (end of filling) is displayed. When the tank is filled switch off the pump (or close
the valve) and press the ENTER key to stop measurement, then enter the measured volume
(100 litres). Validate by pressing the ENTER key: the calculated K factor is displayed.
Please note: The transmitter uses the last either calculated or manually entered K factor.
K.FACTOR
ENTER
Modifying the K
factor automati-
cally resets the
totalizers.
CURRENT
CURRENT
Please note:
To shift the decimal point simultaneously press keys
it is always displayed at the right hand side of the blinking digit. There are two possible
positions: "0000.0" or "000.00".
TEACH NO
ENTER
0......9
ENTER
TEACH Y
ENTER
Start of measurement
ENTER
Display of the calculated K factor
F-33-
8025 / 8035 KONTI-DOS
K =00.00
Enter the sensor K factor (0,01
to 80000).
0......9
To shift the K factor decimal
point.
0......9
K= 117.60
FILL END
ENTER
End of measurement
Enter the measured
volume. Use the same
unit as the totalizer unit.
K= 79.30
ENTER
.
0......9
0000.0 L
0......9
0100.0 L