Fault
Possible Cause
Electrode lost adhesion to skin.
Electrode to dry and loses
Trip off - Elec-
adhesion to skin.
trode symbol
flashing, current
Lead wire not connected
won't increase
properly.
above 8 mA)
Lead wire is faulty.
Device is damaged.
Flat battery or no batteries
installed in device.
Battery installed in wrong
Device does not
orientation.
power on.
Battery connector damaged or
flattened.
Device is damaged.
Batteries almost flat.
Device powers
off during use.
Device is damaged.
Device is in use near a device
Current output is
emitting strong radio or mag-
unstable
netic interference.
Treatment
is painful or
Patients skin has cuts or abra-
uncomfortable
sions in the area of electrode
when using skin
placement.
electrodes.
Treatment
is painful or
Patient reaction to probe
uncomfortable
materials.
when using a
probe.
Skin becomes
Device is being used for
red after treat-
periods which are too long for
ment.
the patient.
Treatment is
uncomfortable,
Intensity too high for required
tingling sensa-
treatment.
tions during use.
Dolito Bedienungsanleitung / Operators Manual

Trouble Shooting Guide

42
Solution
Check electrodes and replace if
necessary.
Replace electrode.
Check lead wire connections and
ensure they are fully inserted.
Replace lead wire.
Return device to manufacturer for
repair.
Check / Replace batteries.
Remove batteries and install accor-
ding to diagrams in battery compart-
ment.
Remove batteries, check metal parts
in battery compartment for damage.
Return to manufacturer for repair.
Check battery icon during use and
replace batteries if low.
Return to manufacturer for repair.
Move at least four meters away from
the offending device and resume
treatment.
Discontinue use until skin is fully
healed.
Discontinue use and consult a doctor
for an allergy check if discomfort
continues.
Use device for a shorter treatment
duration.
Reduce intensity setting until treat-
ment is comfortable.
Note: Tingling sensations are normal
and should be felt.