safe storage and transport. The instruction manual must always be kept with the centrifuge!
Rotor(s) / Accessories will be packed separately.
WARNING: Lifting Hazard. Single person lift could cause injury. Use a
mechanical lifting device or team lifting procedures when lifting or moving
the equipment.
Please refer to section 8.3 for details about lifting the equipment out of the
packaging.

2.1.1 Delivery package

2.2 Selecting the Location

NOTE!
Avoid excessive vibrations, heat sources, air current, or rapid temperature changes.
The centrifuge should be installed on an even, solid and level surface, if possible on a laboratory cabinet
/ table or some other solid vibration free surface.
During centrifugation, the centrifuge must be placed in a way, that there is a minimum space of 30 cm
on each side of the unit according to the standards EN 61010-2-020.
Do not place the centrifuge next to a window or a heater, where it could be exposed to excessive heat,
as the performance of the unit is based on an ambient temperature of 23°C.

2.3 Installation

Follow these steps:
Check whether the power supply corresponds with the one specified on the manufacturer's
rating label, which is located on the rear panel.
For FC2706/FC2516, the power line should be protected by a 10 A rating circuit breaker (type K).
In case of emergency, there must be an emergency switch off installed outside the room in
order to disconnect the power supply from the unit.
Connect the centrifuge to a grounded power receptacle.
Connect the centrifuge with the mains. (The socket for the power cord must be easy to reach for
disconnection)
Turn the instrument on using the mains power switch.
Open the lid by using the Door Open button.
Remove the transport-securing device of the motor.

2.4 Safety precautions during operation and warranty

Do not operate the centrifuge in case it's installed incorrectly.
Do not lean on the centrifuge during operation.
Do not stay within the 30 cm clearance envelope longer than necessary for operational reasons.
Quantity
1
1
1
1
1
Description
Centrifuge FC2706/FC2516
Power Cable
Warranty Card
Instruction Manual / User Guide
Rotor Key
EN-3

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Frontier fc2706Frontier fc2516

Inhaltsverzeichnis