1. Localisation
Fig. 24: Locating the adjustable unit
2. Verifying
proSA Verification Tool
Fig. 25: Place centrally over the valve
tion Tool
WARNING
Failing to centre the Verification Tool over
the valve can lead to erroneous readings!
Fig. 26: Measuring the pressure setting
a) correctly
b) incorrectly
3. Adjustment of the opening pressure
When adjusting the proSA preoperatively
through the packaging, moderate force
with the Adjustment Tool is sufficient. DO
NOT USE THE BUTTON. Strong pressure
can cause damage to the housing, which
might a ect valve function.
Due to postoperative swelling of the skin,
adjusting the valve may be difficult within
the first few days!
Please ensure that the opening pressure is
changed by no more than 8 cmH O per each
adjustment for the adjustable DP unit and
no more than 16 cmH O per each adjust-
ment for the adjustable gravitational unit.
Verifica-
Fig. 27: Rotor rotation during adjustment
a) false 1. non-adjustable range
NOTE
CAUTION
WARNING