4.
Remove the gauge again; it must be possible to pull it out easily!
5.
Replace the pot over the coil and fix it with its three screws (2).
6.
Align the transport lock (7) parallel to the cell chassis using the centring gauge [11] and fasten
it with its two slightly loosened screws (6). The gauge must then be able to be pulled out easily.
The annular gap between the transport safety device and the balance arm must be even (view X
of the weighing cell overview).
7.
Fasten the analog meter [13] to the mounting support using two spacers [15].
8.
Connect the ribbon cable [1] to the meter; the bare contacts towards the top.
9.
Power the meter with the power adapter of the balance.
10. Turn the adjusting screw [12] into the floating sensor body; if necessary, loosen the circuit
board on the magnet pot slightly.