Herunterladen Inhalt Inhalt Diese Seite drucken

HACH LANGE HIAC 9703+ Bedienungsanleitung Seite 16

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
Figure 7 Attach the syringe
1 Plunger bolt
2 Syringe
1. Push the ON/OFF switch to turn the instrument on.
2. Push the LOAD button under the syringe compartment to set the
plunger bolt at the lowest point. Refer to
3. Loosen and remove the plunger bolt from the syringe drive.
4. Hold the syringe and push the bolt through the plunger hole.
5. Push the bolt in the syringe drive hole and turn the bolt by hand to
tighten.
16 English
Figure
7.
3 Syringe drive
4 Valve
6. Pull the syringe barrel up and install the syringe threads in the valve.
7. Turn the syringe by hand to tighten.
Note: Do not overtighten.
Set the sample probe position
The lift mechanism for the sample probe can be set to automatically go to
the correct position for sampling.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis