Measurement
Cuff pressure 0 – 299 mmHg,
range
systolic 40 – 250 mmHg,
diastolic 40 – 250 mmHg,
Pulse 40 –199 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation accord
inaccuracy
ing to clinical testing: systolic 8 mmHg /
diastolic 8 mmHg
Memory
2 x 50 memory spaces
Dimensions
L 156 mm x W 105 mm x H 74,5 mm
Weight
Approx. 315 g (without batteries)
Cuff size
24 to 36 cm
Permissible oper
+10 °C to + 40 °C, 40 – 85 % relative air
ating conditions
humidity (noncondensing)
Permissible stor
10 °C to + 60 °C, 10 – 90 % relative air humid
age conditions
ity, 800 –1050 hPa ambient pressure
Power supply
4 x 1,5 V
Battery life
For approx. 300 measurements, depend
ing on the blood pressure level and/or pump
pressure
Accessories
Instruction for use, 4 x 1.5 V AA batteries,
storage pouch
Classification
Internal supply, IPX0, no AP or APG, continu
ous operation, type BF applied part
Technical information is subject to change without notification
to allow for updates.
All manuals and user guides at all-guides.com
AA batteries
• This unit is in line with European Standard EN 6060112
and is subject to particular precautions with regard to elec
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated Cus
tomer Service address or found at the end of the instructions
for use.
• This device is in line with the EU Medical Devices Directive
93/42/EC, the "Medizinproduktegesetz" (German Medi
cal Devices Act) and the standards EN 10601 (noninva
sive sphygmomanometers, Part 1: General requirements),
EN 10603 (noninvasive sphygmomanometers, Part 3: Sup
plementary requirements for electromechanical blood pres
sure measuring systems) and IEC 80601230 (Medical elec
trical equipment – Part 2 – 30: Particular requirements for the
safety and essential performance of automated noninvasive
blood pressure monitors).
• The accuracy of this blood pressure monitor has been care
fully checked and developed with regard to a long useful life.
If using the device for commercial medical purposes, it must
be regularly tested for accuracy by appropriate means. Pre
cise instructions for checking accuracy may be requested
from the service address.