10. Appendix
10.5 Electromagnetic Compatibility (EMC) Information
Guidance and manufacturer's declaration - electromagnetic emissions
The VEX-P300 is intended for use in the electromagnetic environment
specified below. The customer or the user of the VEX-P300 should assure
that it is used in such an environment.
Emissions test
RF emissions CISPR
11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
84
Compliance
Group 1
Class A
Applicable
Applicable
EzRay Air Portable (Model: VEX-P300) User Manual
Electromagnetic environment -
guidance
The EMISSIONS characteristics of
the VEX-P300 make it suitable for
use in industrial areas and
hospitals (CISPR 11 class A,). If it
is used in a residential
environment (for which CISPR 11
class B is normally required) this
equipment might not offer
adequate protection to
radiofrequency communication
services. The user might need to
take mitigation measures, such as
relocating or re-orienting the
equipment.
The VEX- P300 is suitable for use
in all establishments and may be
used in domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.