Herunterladen Inhalt Inhalt Diese Seite drucken

Controller Controls, Indicators And Connectors - Varian Turbo-V 70 Bedienungshandbuch

Vorschau ausblenden Andere Handbücher für Turbo-V 70:
Inhaltsverzeichnis

Werbung

INSTALLATION
!
Connection to the mains must be made in accordance
with the local law. Always connect the ground wire and
use a properly grounded power socket to avoid
electrical shock. High voltage developed in the
controller can cause severe injury or death. Before
servicing the unit, disconnect the input power cable.
The Turbo-V controller can be used as a bench unit or
a rack module, but it must be positioned so that free air
can flow through the holes. Do not install or use the
controller in an environment exposed to atmospheric
agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk.
During
operation,
the
conditions must be respected:
temperature: from O °C to +40 °C
relative humidity: 0 - 95% (without condensation).
To connect the controller to the pump use the specific
cable supplied with the controller.
The controller model 969-9506 is not equipped with the
pump cable. It can be requested as an accessory;
detailed information is supplied in the paragraph
"Accessories and Spare Parts" of the appendix
"Technical Information". It must be connected between
the controller connector J16 and the pump connector.
The two connectors at the two cable extremity cannot
be connected in a wrong way.
USE
This paragraph describes the fundamental operating
procedures.
Detailed
procedures that involve optional connections or options
are supplied in the paragraph "USE" of the appendix
"Technical Information". Some procedures can be
executed with controller model 969-9506 only when the
accessory "Hand held terminal" is available. Make all
vacuum manifold and electrical connections and refer
to Turbo-V pump instruction manual prior to operating
the Turbo-V controller.
WARNING!
NOTE
following
environmental
NOTE
information
and
operating
To avoid injury to personnel and damage to the
equipment, if the pump is laying on a table make sure it
is steady. Never operate the Turbo-V pump if the pump
inlet is not connected to the system or blanked off.

Controller controls, indicators and connectors

The following paragraph illustrates the hand held
terminal control panel and interconnection panel.
More details are contained in the appendix "Technical
Information".
1. Keyboard
push-button
selection. It is active only when the front panel
operation has been selected. Pressed once, the pump
runs at about 2/3 of the nominal speed. To unselect
the mode, press the push-button again.
2. Keyboard push-button for START, STOP, RESET
mode selection. It is active only when the front panel
operation has been selected. By pressing once the
starting phase begins; if pressed again it stops the
pump. If the pump has been stopped automatically by
a fault, this push-button must be pressed once to reset
the controller and a second time to restart the pump.
3. Keyboard push-button to recall on the display the
cycle number, cycle time and pumplife.
4. Keyboard push-button to recall on the display the
pump current, pump temperature, pump power and
rotational speed. It is always active regardless of the
operating mode selected. Push-buttons 3 and 4, if
pressed together for
controller in a routine where it is possible to program
some operation parameters.
5. LCD back-lighted alphanumeric display: dot matrix 2
lines x 16 characters.
Hand Held Terminal Control Panel
46
INSTRUCTIONS FOR USE
!
WARNING!
for
LOW
SPEED
mode
at least 2 seconds put the
87-900-870-01 (B)

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

969-9506

Inhaltsverzeichnis