Herunterladen Inhalt Inhalt Diese Seite drucken

Promag 5D4C, 5L4C, 5W4C Transmitter, Remote Version - Endress+Hauser Promag 400 Index C Installationsanleitung

Austausch anschlussplatine elektronik kompakt und getrennt
Vorschau ausblenden Andere Handbücher für Promag 400 Index C:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EA01098D
8
Promag 5D4C, 5L4C, 5W4C transmitter, remote
version
8.1
Replacing the connection board
Proceed as described in Section 7.4→  8.
8.2
Reassembling the transmitter housing, remote version
8.3
Standard transmitter
NOTICE
Poor measuring performance!
If the grounding switch is in the wrong position, this impacts the measuring performance!
Check that the grounding switch is in the correct position and correct if necessary!
1 →
Version A: If a grounding switch
is provided on the ISEM: set the
grounding switch of the ISEM to
the "right" position (open).
12
2
Version B: If a jumper is
provided on the ISEM: plug in
the jumper for the ISEM
grounding setting on the outside
(open).
Endress+Hauser

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Prosonic flow 400Promag w 400

Inhaltsverzeichnis