8.2 Restore factory setting

a.
1 x
> 3 s
b.
c.
1 x
The instrument is permanently adjusted for aqueous solutions but can also be adjusted for solutions
of different density, viscosity, and temperature. The Transferpette® electronic can be adjusted in any
mode (except GEL mode).
Transferpette® electronic 997209
e.
Confirm volume: Press the Enter key.
ð
The display shows the tested and corrected volume.
ð
The CAL symbol is continuously displayed to confirm
that an adjustment has been made.
The CAL symbol permanently shown in the display indicates
that an adjustment has been made.
a.
Enter CAL mode: Pressing and holding (> 3 s) the menu
key invokes the CAL mode.
ð
The display shows 'on'.
ð
'CAL' flashes.
b.
Switch off CAL mode: Pressing one of the arrow keys de-
activates the CAL mode.
ð
The display changes from 'on' to 'off'.
ð
'CAL' continues to flash.
c.
Restore factory condition: Press 'Enter' key.
ð
The permanently displayed CAL icon has disappeared.
ð
The instrument has been returned to the original fac-
tory condition.
For the Transferpette® electronic , a volume offset is applied
during adjustment (i.e. the volume changes by the same
amount over the entire volume range of the pipette). It is rec-
ommended to perform the adjustment at 50% of the nominal
volume.
NOTICE
8 Adjustment – Easy Calibration