4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider must ensure that all per-
sonnel handling the product are provided with the operating instructions & performance data.
Do not use the product without properly familiarization of the product and its safe operation
as defi ned in this Instruction for use. Ensure user is aware of particular information and knowl-
edge required for the gas in use.

5. PRODUCT DESCRIPTION

5.1. MM90 STANDBY WITH PRESSURE SWITCHES
The MM90 STANDBY manifold covered by these instructions is designed to be used as a third
source of supply in medical central gas systems. The manifold will deliver gas when the nominal
supply system pressure falls below a set level. This is a back up source.
The manifold consists of manifold pressure regulator A , inlet shut off valve B and purge valve
C. Pressure gauge D is fi tted to the inlet side downstream of B. It shows the content pressure of
the cylinder bank. An extra gauge E is also fi tted to outlet F to show the outlet pressure. Pres-
sure switch no. 1 is fi tted to the inlet downstream of B to indicate empty cylinders . Pressure
switches fi tted to the outlet F indicate high no. 2 ("H" marking) and low no. 3 ("L" marking) outlet
pressures. The pressure switches shall be connected to an alarm panel (Gas alarm) or central
monitoring system to give a visual and audible indication of the condition of the manifold.
Fig. 1