Herunterladen Inhalt Inhalt Diese Seite drucken

Medela Mini Electric Gebrauchsanweisung Seite 163

Vorschau ausblenden Andere Handbücher für Mini Electric:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 30
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
Electromagnetic immunity
The electric breastpump Mini Electric is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Mini Electric should
assure that it is used in such an environment.
IEC 60601
Immunity test
test level
3 Vrms
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
Compliance
Electromagnetic
level
environment – guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the electric
breastpump Mini Electric, including
cables, than the recommended
separation distance calculated from
the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
d = 0,35√P
d = 0,35√P
10 Vrms
d = 0,7√P
where P is the maximum output power
10 V/m
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended sepa-
ration distance in metres (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey,
than the compliance level in each fre-
quency range.
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
should be less
b
163

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis