With the entry of such a waiting time, the acceptance/recording of a test point is
blocked for this duration. (In the above example, after the first test point has been
recorded, 30 seconds must be waited before the second test point can be
recorded.)
Should already stored test points be deleted, because the new
calibration has less test points than the previous one, the "CLEAR"
button has to be pressed. This will delete the current and all follow-
ing test points. (The erasing procedure will take some seconds.)
Note: For deleting the complete stored calibration data all at once
please see Chapter 10.7.
38
Optional DKD waiting time [sec]
Input via numeric keypad & confirma-
tion with
ENTER
WIKA Operating Instructions ProcessCalibrator