Electromagnetic immunity

Electrostatic discharge
Test
standard
EN 61000-4-2
Radiated RF electromagnetic field
Test
standard
EN 61000-4-3
The product must not be affected by electrical discharge
between a person and the product in such a way as to
initiate actions or render components defective.
Immunity test level
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
The product must not be affected by radiating RF signals,
for example from a mobile phone or radio, that could
cause the product to take action or to make components
defective.
Immunity test level
15 V/m 80 MHz to 1000 MHz (only on the front)
15 V/m 1,01 GHz to 3 GHz
80 % AM 1 kHz