Menu options – Calibration Mode
GB
If the instrument is not able to communicate with the IRiM, a timeout
E 132
occurs after approx. 2 minutes. The error E 132 is displayed for approx.
4 seconds. Subsequently, the instrument switches to test mode (see
also IRiM manual).
Store
Date
2
2
3
Time
Cal
After confirming the selection with the [MODE] key the value to be
Mode
edited will be shown for 2 sec.
The setting starts with the year (YYYY) followed by the actual value
SET
to be edited. The same applies for month (mm), day (dd), hour (hh)
DATE
and minutes (mm). Set the minutes first in steps of 10, press the [!]
key to continue setting the minutes in steps of 1.
YYYY
(2 sec.)
Increase the value by pressing the [MODE] key.
Mode
Zero
Decrease the value by pressing [ZERO/TEST] key.
Test
!
Proceed to the next value to be edited by pressing [!] key.
After setting the minutes and pressing the [!] key the display will show
"IS SET" and the instrument returns to the measurement mode.
Store
Date
4
Cal
4
Time
Cal
Note:
user calibration (Display in calibration mode)
cAL
factory calibration (Display in calibration mode)
CAL
After confirming the selection with the [MODE] key the instrument will
CAL
show CAL/PtCo.
PtCo
Fill a clean vial with filtrated deionised water, screw the cap on and
place the vial in the sample chamber making sure that the
are aligned.
Zero
Press the [ZERO/TEST] key.
Test
The "Range" symbol flashes for approx. 8 seconds.
PtCo
0.0.0
The display shows the following in alternating mode:
CAL
Perform measurement with a standard of known concentration as
described.
Zero
Press the [ZERO/TEST] key.
Test
PtCo
The "Range" symbol flashes for approx. 3 seconds.
MD100_4 10/2014
3
Setting date and time (24-hour-format)
User calibration
marks