Appendix (cont.)
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 6)
212
Recommended separation distances between portable and mobile RF
communications equipment and the Symphony electric breast pump
The Symphony breast pump is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Symphony
electric breast pump can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Symphony electric breast pump as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
to the frequency of the transmitter, where
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
d = 0.35√P
0.04
0.11
0.35
1.11
3.50
in metres (m) can be estimated using the equation applicable
d
End of document
80 MHz to
800 MHz
d = 1.2√P
0.12
0.37
1.2
3.7
12
is the maximum output power rating of the
P
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.74
2.3
7.4
23