Safety instructions
6.6.1 IECEx
The approved digital sensors with Memosens technology may be connected via the following measuring
cable
1.
CYK10
2.
or a structurally identical and in hardware and function identical and Memosens measuring cable
only to the IECEx approved intrinsic safe sensor output circuits of Transmitter Liquiline M CM42 or
alternatively to an IECEx certified, intrinsic safe Memosens sensor output providing the following maxi-
mum values as described below. In particular the effective inner inductivity and capacity of the appro-
ved, intrinsic safe sensor output must not exceed the values given below:
1. Entity Parameters
U
= 5,1 V
0
I
= 130 mA
0
P
= 166 mW (linear output characteristic)
0
= 15 μF
C
i
= 95 μΗ
L
i
22
All manuals and user guides at all-guides.com
2. Entity Parameters
U
= 5,04 V
0
I
= 80 mA
0
P
= 112 mW (trapezoide output characteristic)
0
= 14,1 μF
C
i
= 237,2 μΗ
L
i
CYK10/11