Defender 6000 Indicators

Dynamic Mode

3.5
Use this application to weigh an unstable load, such as a moving animal.

3.5.1 Enter the Mode

To enter the Dynamic Weighing Mode from any application mode:
1.
Press and hold the Mode button until
2.
The display shows

3.5.2 Start Dynamic Weighing

1.
To start:
 When operation type is manual. Place the load (more than or equal to 5d) on the pan and press the
Function button to start the averaging process.
 When operation type is semi-automatic/automatic. Place the load (more than or equal to the Start
Weight) on the pan, and the terminal will start averaging process automatically
Note: The display must be at zero gross or net value before placing the load on the pan.
2.
During the averaging period, the countdown timer decreases in one second increments (For example, the
set average time is 5s).
Note: If the set average time is 0s, the countdown timer is not displayed.
3.
The readings are averaged and held on the display when the countdown has completed in both
Countdown and Continuous mode.
In addition, for DT61XWE model:
 The tilde symbol will blink indicating that the current weight is being held in Countdown Mode.
 The tilde symbol will blink indicating that the current weight is being averaged in real time in Continuous
Mode.
Note: If the set average time is 0s, the first weight larger than 5d will be displayed and hold.
4.
To reset the countdown timer:
 When the operation type is manual/semi-automatic, press the Function button to reset the countdown
timer when the countdown is running. The display shows
 When operation type is automatic, remove the load from the pan, and the average weight will still be
displayed until the duration time is over. Then the display shows
Note:
rEADY
The
Please refer to Check in the Weighing Mode section for how to use Check in the Dynamic
Weighing mode.

3.5.3 Application Settings

The application can be customized for user preferences. Please refer to Application Settings section in
Weighing Mode for details about how to enter application settings.
The Dynamic Weighing Configurations are defined below (defaults in Bold).
Item
Available Settings
Dynamic Mode
Countdown (
dyNAmM
Continues (
Off (
Dynamic Operation
Manul (
d.tyPE
Type (
)
Semi-auto (
Auto (
Start Weight
0 ~ Capacity Weight
S.wWGt
(
)
Duration Time
1 ~ 10 s
d.tImME
(
)
Average Time
0 ~ 30 s
A.tImME
(
)
dYNAmM
rEADY
.
display must be at zero gross or net value in order to reset the countdown timer.
C.DOwWN
)/
CONt
)/
OFF
)
mMAN
)/
semMI
)/
AUtO
)
is displayed.
rEAdY
, and start to re-count.
Comments
Count down:
There is a countdown time.
Continuous:
Averaging will be continued after the countdown time.
Manual:
The averaging process is started and reset manually.
Semi-auto:
The averaging process is started automatically and reset
manually.
Auto:
The averaging process is started and reset automatically.
Dynamic weighing will start when the load is bigger than the
start weight (for Semi-auto and Auto mode).
It is the time for the display to remain the dynamic weighing
result after the load is removed.
Time in Seconds.
If the average time is 0, the first stable weight (more than or
equal to 5d) will be the result.
rEAdY
, and start to re-count.
EN-21