STYLISTIC ST6012 MED
Table 3: Guidelines and manufacturer's declaration – Electromagnetic
The STYLISTIC is intended for operation in the electromagnetic environment specified below. The
customer or user of the STYLISTIC should ensure that it is used in such an environment.
Immunity
IEC 60601
testing
Test level
Conducted RF
3 V
eff
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
6 - English
immunity
Level of
Electromagnetic environment - guidelines
compliance
Portable and mobile RF communications
equipment should be used no closer to any part of
the STYLISTIC, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
3 V with
d
docking unit
and/or
peripheral
cabling, but
without analog
modem
3 V/m
d
80 to 1000 MHz
3 V/m
1000 to
2500 MHz
d
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
3,5
P
V
2
3,5
80 MHz to 800 MHz
P
E
1
7
800 MHz to 2.5 GHz
P
E
1
c
A26391-K279-Z290-1-8N19, edition 1