Note
The accuracy of the 10A range is valid from 20% to 100% of the range.
DC current
Range
4A
10A
-Overload protection: 10A/690V fast-blow fuse protection
-Maximum measured current: 10A (for measured current greater than 2A:
Duration < 10s, measurement interval > 15min)
AC current
Range
4A
10A
-Overload protection: 10A/690V fast-blow fuse protection
-Frequency range: 40Hz-400Hz
-Display: sine wave rms (average response)
-Maximum measured current: 10A (for measured current greater than 2A:
duration < 10s, measurement interval > 15min)
Note
The accuracy of the 10A range is valid from 20% to 100% of the range.
Resistance
Range
400Ω
4kΩ
40kΩ
400kΩ
4MΩ
40MΩ
03005 / 03007
Resolution
0.001A
0.01A
Resolution
0.001A
0.01A
Resolution
0.1Ω
0.001kΩ
0.01kΩ
0.1kΩ
0.001MΩ
0.01MΩ
-27-
EN
DE
RU
KO
Accuracy
±(1.0%+7)
±(1.5%+7)
Accuracy
±(1.5%+7)
±(2.5%+7)
Accuracy
±(1.0%+5)
±(0.5%+5)
±(1.5%+5)
±(3.0%+5)
PT