Troubleshooting

Trouble
- Tip dripping (instrument leaks)
- Aspirated volume too low
- Delivered volume too low
- Slow suction or none
- Volume too high
GA Transferpette Einkanal BRANDT22 22
GA Transferpette Einkanal BRANDT22 22
- 22 -
Possible Causes
- Unsuitable tip
- Tip not seated tightly
- Pipette shaft cone loose
-Seal contaminated
- Seal damaged
- Non-aqueous solutions pipetted
- Piston damaged
- Shaft clogged
- Pipetting key pressed too far (into
the blow-out position) before
sample take up.
Action to be taken
- Only use high-quality tips
- Press tip on fi rmly
- Tighten clockwise see "Handling",
page 6
see "Cleaning", page 16
see "Specifi cations/Spares", page 25
see "Adjustment", page 11
see "Spares", page 25
- Clean shaft see "Cleaning", page 16
- Operate properly
see "Pipetting", page 8