If a leakage is detected again, apply the "After use" procedure below and
return the valve for service or repair.
6.3. OPERATIONS
•
Set the suction ejector ON by pulling up the integrated handwheel.
•
Lock the patient side inlet with your finger.
EN
•
Set the vacuum level to the needed level for your therapy by turning the
integrated handwheel in the anti-clockwise direction and checking the
vacuum indicator (gauge).
If vacuum value is reduced in the absence of flow, the vacuum indicator
remains at maximum. See the description of the non-return valve.
•
The suction ejector is ready to be used.
6.4. AFTER USE
•
Before disconnecting ejector from the medical pipe line system terminal
unit/mobile gas source:
•
•
Follow the associated equipment manual. For example: If the collecting
container is overfilled, extracted fluids are sprayed on / into the patient
when the vacuum source is abruptly interrupted.
7. ACCESSORIES
Not available.
For closer information about the possible optional equipment refer to the
section 8.2 associated equipment.
8. CLEANING
8.1. SUCTION EJECTOR
MediEject II must be cleaned and disinfected externally after every pa-
tient.
If there is risk that suction ejector is exposed to contamination following
cleaning shall be used:
Remove general contamination with a soft cloth damped in oil free oxygen
compatible soap water and clean with deionised water.
Disinfection can be done with an alcohol-based preparation (moistened
wipe). If another cleaning agent is used, check that it is not abrasive and is
compatible with the product materials (including labels) and gas (a suitable
cleaning agent - e.g. Meliseptol).
turn it OFF, by pushing down the integrated handwheel and set
vacuum level to minimum by turning the integrated handwheel
clockwise
remove suction tubing connection from patient side
8/176