Safety information
You have decided on an appliance that offers you a high degree of safety. In order to ensure safe and correct application, before
using initially it is necessary to fully read these operating instructions.
The following safety precautions are applicable:
Immediately before using, carry out a function check of the voltage
|
tester (VDE regulation 0105, Part 1). Make sure that the test line and
appliance are fully serviceable. Check the appliance on a known
source of voltage, e.g. 230 V socket.
|
Hereby, if the display fails to indicate one or more functions, do not use
the appliance and it must be checked by a specialist.
|
Only hold the instrument on the handles (3) below the mechanical
mark (2), avoid contact with the test probes (1)! (Refer to the figure)
Only carry out for isolation from the power supply on both poles!
|
|
The appliance must not be operated in a moist environment!
|
Do not use with the battery compartment open! Whilst replacing the
battery, the test probes must be removed from the test circuit.
Correct indication is ensured in the temperature range of -15 °C to
|
+45 °C.
Always keep the appliance dry and clean. Only use a moist cloth to
|
clean the housing.
The additional warning symbol and vibration for voltages > 35 V are
|
only for warning of hazardous voltages, not for test purposes!
Immediately replace the batteries if a warning signal sounds shortly af-
|
ter switching on the instrument, or when the instrument is switched on
and immediately switches off.
The different signals of the voltage tester (including the ELV limit value)
|
displayed must not be used for measurement.
General
Voltages have priority. If no voltage is applied at the test probes (< 4 V), the appliance is in the continuous test mode.
Function
To switch on the instrument, simply hold the test probes on one another.
In stand-by mode, the orange "Rx /" LED illuminates.
Attaching and removing the adapter for the test probes makes it easier to carry out tests at the socket.
For safe keeping, a suitable bracket is located at the captive test probe protection.
The nominal In current is a max imum of 3.5 mA.
Self-Test / Auto Power-Off
When initially switching on, either when inserting a new battery or when the test probes are held on one another if the instr ument is
off (the "Rx /"-" LED does not illuminate), the instrument illuminates all optical displays (LED's). Shortly after, the instrument
changes to the continuity check.
Hereby, if the display fails to indicate one or more functions, do not use the appliance and it must be checked by a speciali st.
To test, in order to ensure that the instrument is ready to use, place the test probes on one another. This should be carried out im-
mediately before and after use. The green "Rx /" LED must clearly illuminate.
If the LED does not, or only faintly illuminates, replace the batteries.
If the appliance does not function with new batteries, i t must be protected from being used incorrectly.
After a period without using, the instrument automatically switches off by the "Auto Power Off" function.
Thereby, the orange "Rx /" LED goes out. To switch on the instrument again/self test, momentarily hol d the test probes on one
another.
®