14. Troubleshooting

Review the information in the table below to troubleshoot operating problems.
PROBLEM
I. Optical Section:
LED operates, but field of view re-
mains dark.
Field of view is obscured or not evenly
illuminated
Dirt or dust is visible in the field of
view.
The image (using the microscope in
the bright field) appears doubled
Visibility is poor.
Image is poor.
Contrast is poor.
Details are indistinct.
One side of the image is unfocused
II. Mechanical Section:
Coarse focus knob is hard to turn
Focus is unstable
III. Electrical Section
LED doesn't turn on.
Brightness is not enough
Light blinks
CAUSE
Power supply is unplugged.
Brightness is too low
Revolving nosepiece is not correctly
engaged.
Oil is not present or not in sufficient
quantity on the front lens of the con-
denser and on the slide
Dirt/dust on the specimen
Dirt/dust on the eyepieces
Aperture iris diaphragm is stopped
down too far.
The condenser is not well centered or
it is in a wrong height
Revolving nosepiece is in an incorrect
position
Aperture iris diaphragm (when work-
ing in brightfield) is too closed or too
open.
Dust or dirt on lenses (condenser,
objectives, eyepieces and slide)
For transmitted light observation, the
coverglass thickness must not exceed
0.17mm
Focus is not even
A sample not suitable for dark field
observation is being used.
Revolving nosepiece is in an incorrect
position
Slide is mounted not in a flat position
(tilted)
Poor quality of the glass slide
Tension adjustment ring is too tight
Tension adjustment ring is too loose
Power supply not connected
Brightness setting is too low
Power supply not well connected
Page 23
SOLUTION
Connect
Set brightness to a proper level
Make sure that the revolving nose-
piece clicks properly into place.
Check for the correct presence of oil
Clean the specimen
Clean the eyepieces
Open aperture iris diaphragm.
Set the condenser according to Koe-
hler settings.
Move the nosepiece to a click stop
Adjust aperture iris diaphragm.
Clean thoroughly.
Use a coverglass with thickness
0.17mm
Slide holder is not flat. Move the spe-
cimen to a flat position.
Use a suitable sample
Move the nosepiece to a click stop
Place the specimen in a flat position
on the stage
Use a glass slide with higher quality
Loosen tension adjustment ring
Tighten tension adjustment ring
Check for proper connection
Adjust brightness
Check for proper connection