8.4.2 Adjustment test with internal weight

During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press the [TARE] key. "CAL" is displayed.
Press the [TARE] key. "E CAL" is displayed.
Repeatedly press the [CAL] key until "i tESt" appears.
Press the [TARE] key. The display changes automatically in this
order: "tESt 2", "tESt 1", "tESt 0"
After this, the difference to the previous adjustment will be dis-
played for several seconds.
After "tESt End" was displayed, the balance will automatically
return to weighing mode.