CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
► Use caution when working with the product.
► If the product is damaged, check it for proper function and readi­
ness for use.
► In case of changes in or loss of functionality, do not continue
using the product (see "Signs of changes in or loss of functional­
ity during use" in this section).
► Take any necessary measures (e.g. repair, replacement, inspec­
tion by the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an
altered gait pattern, a change in the positioning of the prosthetic com­
ponents relative to each other and by the development of noises.

4 Scope of delivery

Quantity
Designation
1
Instructions for use
1
Adapter
Only for 2R41=*, 2R48 tube adapter
4
Set screw
Spare parts/accessories (not included in the scope of deliv­
Quantity
Designation
Special adhesive with hardener
Single component pack for 2R40=1 con­
sisting of: 1 cap screw, 1 flat washer
Reference num­
ber
506G3=M6x10
ery)
Reference number
636W28
636W18 and 636W19
2D6=M8
Spare parts/accessories (not included in the scope of deliv­
Quantity
Designation
Single component pack for 2R40=2 con­
sisting of: 1 cap screw, 1 flat washer

5 Preparing the product for use

CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connec­
tions
► Clean the threads before every installation.
► Apply the specified torque values.
► Follow the instructions regarding the length of the screws and
about how to secure the screws.
5.1 Installing the foot adapter
Required tools and materials: Grinding tool, 710D20 torque
>
wrench, degreasing cleaning agent (e.g. 634A3 acetone), special
adhesive
and
hardener
parts/accessories)
1) For prosthetic feet with wooden components, roughen the con­
nection surface and remove the grinding dust.
2) INFORMATION: The foot adapter has to be glued to the
prosthetic foot. Use the specified special adhesive for this
purpose.
Mix the special adhesive and hardener well together and apply to
the connection surface.
ery)