Hei-VAP Precision
Error / message
The heating bath is
completely evaporated.
Please refill the heating bath
medium.
An illogical temperature
difference was measured
between the heating bath
sensors. The temperature
measurement is defective.
The temperature sensor for
the heating bath control is
defective.
(hardware monitoring unit)
The temperature sensor for
the heating bath control unit
is broken or bath medium
that is too hot was filled in.
The temperature sensor for
the heating bath control unit
is defective or bath medium
that is too cold was filled in.
The lift motor is defective or
blocked.
The set safety pressure was
exceeded.
The vacuum box is not
connected or defective.
The ventilation valve is
defective.
Downloaded from
www.Manualslib.com
Troubleshootimg
manuals search engine
Cause
Heating bath is
completely evaporated
Heating bath sensor
variance > 10 K
▪ Sensor monitoring
via hardware
▪ Heating bath sensor
is broken or short
circuited
▪ Heating bath sensor
is broken
▪ Heating bath is more
than 217 °C
▪ Heating bath sensor
short circuit
▪ Heating bath is below
0 °C
Lift end stop switch is
not reached after 10
seconds
Pressure sensor safety
pressure is exceeded
Vacuum box is not
connected or defective
Vacuum valve short
circuit
45
Remedy
➜ Deactivate the unit
➜ Check if the maximum
temperature limiter has
responded. If yes, reset
(see chapter "Cleaning
and maintenance,
Maintenance")
➜ Fill in heat transfer fluid
➜ Contact service
➜ Contact service
➜ Switch off device
➜ Cool down heating bath
fluid
➜ Contact service, if no
heating bath fluid was
filled in
➜ Switch off units
➜ Cool down heating bath
fluid
➜ Contact service if heating
bath fluid was not filled in
➜ Contact service
➜ Deactivate the unit
➜ Check if the pressure
sensor has responded. If
yes, reset
➜ Contact service
➜ Check the connection
between the vacuum box
and the evaporator.
➜ Contact service
➜ Check connection
between vacuum box and
evaporator
➜ Contact service