Herunterladen Inhalt Inhalt Diese Seite drucken

Reusable Pump Set - Before First Use, Once A Day And Always Before Passing Onto The Next Patient - Medela Lactina Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Lactina:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

5. Cleaning
Warning
16
l Use by more than one mother without
adequate sterilisation may pose a health
risk.
l All parts must be cleaned after every
use. They must additionally be disin-
fected and sterilized (e.g. autoclaved)
once a day and always before being
passed onto the next mother.
Caution
l Only use drinking-quality water for
cleaning.
l for chapter 5.4 Cleaning the breast-
pump: Unplug the breastpump before
cleaning.
Note
l for 5.5 Sterilisation: Only sterilise clean
and intact parts. Do not stack parts on
top of one another. To avoid damage,
do not apply external pressure.
i
Info
l These cleaning instructions are general
recommendations and may individually
be adjusted based on the hospital's
specific directives and cleaning practices.
5.2 Reusable pump set – before first
use, once a day and always before
passing onto the next patient
1
Disassemble the pump set
into individual parts.
3
134 C
Sterilise all the parts in the
autoclave at 134 °C for three
minutes.

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Lactina selectLactina electric plus

Inhaltsverzeichnis