3.
Expose the monitor to the solution according to the conditions provided in
Ensure that the monitor is exposed for the full duration of the exposure period.
4.
Rinse the monitor according to the conditions stated in
5.
Allow the monitor to air dry.
6.
Store the monitor in a clean environment.
PROCEDURE 3. CLEAN THE VIDEO BATON
Clean and disinfect the video baton after each use.
IMPORTANT
to clean the video baton. The window that protects the camera and light can be scratched,
permanently damaging the device.
This component is heat‑sensitive, and exposing it to temperatures in excess of 60°C (140°F)
will cause damage to the electronics.
Bleach may be used on the video baton, but pay special attention to the stainless steel on the
component, as bleach can corrode stainless steel..
Table 6.
Cleaning Solutions for the Video Baton 2.0 Large
Product
Enzymatic
debridement
agent/detergent
STERIS
®
Prolystica 2x
Concentrate
Enzymatic
Presoak
and Cleaner
SaniCloth
AF3
®
Germicidal
Wipes
1.
Wash the video baton manually using a hospital‑grade equipment detergent or an
enzymatic debridement agent/detergent to remove all foreign material (e.g., soil
and organic material) from the surface of the device. For more information, see
Table
6.
2.
Rinse the video baton in clean, running water.
The video baton can now be disinfected.
16
Do not use metal or abrasive brushes, scrub pads, or rigid tools
As per chemical manufacturer's instructions
Up to 2000 cycles as per the following instructions:
Exposure: Prepare solution in warm water at ⅛ to ½ fl. oz. per gallon
(1–4 mL/L). Soak component for at least 3 minutes. Before removing
from solution, brush all surfaces using a soft‑bristled brush, paying
special attention to hard‑to‑reach areas. Use a cotton swab for the
camera window to avoid damaging the window.
Rinse: Rinse for 3 minutes under warm running water. If component
is soaked for longer than 3 minutes, increase rinse time in proportion
to soak time.
Up to 2000 cycles as per chemical manufacturer's instructions