8
Electrical connection
The following can be connected for measurement signal conditioning:
S Carrier-frequency amplifier
S DC voltage amplifier
designed for strain gauge measurement systems.
The S9M force transducer is delivered with six-wire configuration.
8.1
Connection in a six-wire configuration
When transducers designed with a six-wire configuration are connected to
amplifiers with a four-wire configuration, the sense leads of the transducer
must be connected to the corresponding supply leads: marking (+) with (+) and
marking (-) with (-).
Fig. 8.1
Pin assignment of S9M in six-wire configuration
With this cable assignment, the output voltage at the measuring amplifier is
positive in the pressure direction when the transducer is loaded.
The transducers are delivered with a 7.6 m cable with free ends as standard.
The connection cable shield is connected to the transducer housing. The tr
ansducers with free cable ends must be fitted with CEstandard plugs and the
shielding connected so they cover a wide area. With other connection tech
S9M
(gray)
(black)
(white)
(blue)
(green)
(red)
(-)
A03019_05_YI0_04 HBM: public
Electrical connection
Sense lead (*)
Current feed (*)
Signal ())
Current feed ())
Sense lead ())
Signal (*)
Shield/drain wire, connected to enclosure
ground
21