Vent
Hose ring
Electrolyte 566058 (ECL1)
12
Risk of damaging the membrane due to vacuum in the membrane cap.
Before filling up the membrane cap, uncover the vent by lowering the
hose ring, to allow venting.
On delivery, the membrane cap is not tightly screwed on the shaft.
• Do not touch the reference electrode and the working electrode.
• Do not polish the reference electrode.
• Do not remove the brown or grey coating of the reference electrode.
Unscrew the membrane cap from the shaft and place it onto a clean
base.
For sensors with an internal holder, remove the internal holder from its
package and place the membrane cap and the internal holder separately
onto a clean and not absorptive base.
Prepare the applicable electrolyte.
Carefully fill up the membrane cap to the edge with the enclosed
electrolyte to avoid air bubbles to enter in the membrane cap (Oxygen
can have an unwanted effect on the measurement signal).
english
Instructions
Type 8232
Installationandwiring
566052 & 565164
Sensor
566052
Electrolyte
Electrolyte
566059
566060
(ECS2.1/GEL)
(ECN1/GEL)
Vent
Hose ring
Sensor
565164

Werbung