Functional Principle, Technical Data
2.3
Controller type
Resolution static
Resolution dynamic
Limit frequency
Limit frequency adjustable
Linearity
Max. sensitivity deviation
Long term stability
Synchronous operation
Insulator measurement
Temperature stability
Operating temperature
Storage temperature
Supply
Output
Sensors
Electromagnetic compatibility (EMC)
FSO = Full Scale Output
capaNCDT 6100
Technical Data
DT6100
0.01 % FSO
0.015 % d.M. (2 kHz)
10 Hz / 2 kHz
±0.3 % d.M. (all sensors interchangeable without calibration)
Option LC: ±0.1 % FSO (tuned to one sensor)
±0.1 % FSO
≤ 0.05 % FSO / Month
±0.03 % FSO / °C
Controller: +10 ... +60 °C (+50 to +140 °F)
Sensor: -50 ... +150 °C (-58 to +302 °F)
-10 ... +75 °C (+14 ... +167 °F)
24 VDC / 85 mA (9 ... 36 VDC)
optional ±15 VDC / 85 mA (9 ... 36 VDC)
0 ... 10 V (resistance min. 1.2 kΩ / capacitance max. 1 nF)
optional: 4 ... 20 mA (load max. 400 Ω)
all sensors außer CS005
DIN EN 61326-1: 2006-10 and DIN 61326-2-3: 2007-05
2 kHz
yes
no
Page 40