Herunterladen Diese Seite drucken

Controller Specifications - Agilent Technologies MicroVac 929-0200 Bedienungshandbuch

Werbung

5
Technical Information
MicroVac Controller Description

Controller Specifications

Tab. 1
Operation
Input:
Voltage
Frequency
Power
Output:
Voltage
Current
Power
Current Recorder Output:
- Log Rec Out (see attached
graphs)
Voltage Recorder Output:
- Rec Out
Operating temperature
Storage temperature
Fuse (mains)
Radio interference suppression
Safety
Installation category
Pollution degree
Output Connector:
- High voltage
- Accessory
Cable
Weight:
52/72
Mini, 2 l/s pumps
100, 120, 220, 240 Vac ±10 %, 1-phase
47 to 63 Hz
40 VA maximum
3.300 Vdc nominal ±10 %,
1 mA (short circuit)
1.2 W maximum
0 to 10 Vdc (10 V correspond to 1 mA; 10 nA resolution
0 to 3.3 V corresponding to 0 to 3.3 KV with 1 V corresponding
to 1 KV in linear mode)
Available at pins 4 and 8 of rear panel Accessory connector.
0 °C to +45 °C
-20 °C to +70 °C
2 x T 1A (slow blow) disregarding the mains
EN 55011 class A group 1
EN 61000-4 -2/3/4
EN 61010 - 1
II
2
KINGS type: P/N 10664-I for Model 929-0200 and 929-0201
J 101 input/output signal (9-pin "D" type socket)
Mains, 3 meters long
1.7 Kg (4 lbs)
MicroVac Controller User Manual / 87-900-089-01 (G)

Werbung

loading

Diese Anleitung auch für:

Microvac 929-0201