Operation

3.2 Checking the weighing platform

Before starting a weighing series, perform a function check of the weighing platform and
the connected weighing terminal.
Function check
1. Make sure, that the load plate is closed and locked correctly, see section 4.4 or 4.5.
2. Make sure that the weighing platform is connected to a weighing terminal and that the
weighing terminal is switched on.
3. Make sure that the weighing platform is unloaded and the display of the weighing
terminal shows 0.
4. Load the weighing platform.
The indicator must show a value different from 0.
5. Unload the weighing platform.
The indicator must return to 0.
Verification test
For a verification test refer to the user manual of the connected weighing terminal.
If the verification seal is broken, verification is no longer valid.

3.3 Checking weighing platforms in a pit

➜ Make sure that the weighing platform does not touch the pit frame.
➜ Make sure that the gap (1) between weighing platform and pit frame is free from dirt
deposits.
1

3.4 Checking ramps

➜ Make sure that the weighing platform does not touch the ramp.
➜ Make sure that the gap (1) between weighing platform and ramp is free from dirt
deposits.