Technical Information

Air Cooling Kit Installation

Air Cooling Kit Installation

An air cooling kit (mod. 969-9290) is available for cooling the pump
during heavy operational conditions and whenever the natural air
convection is not sufficient.
Figure 19
Fan specifications:
The fan bracket is shaped so that it can be mounted close to the
pump and in different positions. To fix the fan to the Turbo-V 81-T
case execute the following procedure (see the following figure):
1
2
3
208/228
air flow: 12.8 l/s (27.1 CFM)
input voltage: 24 Vdc
dimensions: 60 x 60 x 25 mm
power: 2.60 W
Fix the fan to the suitable bracket by means of the furnished
screws;
Fix the bracket to the pump body;
Connect the fan supply to the P4 connector of the controller.
Turbo-V 81-T User Manual / 87-900-984-01