InPro 6850i / 6900i / 6950i O
– For correct calibration, a minimum flow rate of the
calibration medium is necessary.
– Make sure that all other parameters, such as
temperature and pressure, are constant.
For continuous applications, we recommend period-
ic recalibration in line with your requirements on
accuracy, the type of process in operation and
your own experience. The frequency of the need for
re-calibration depends very much on the specific
application, and therefore appropriate intervals can-
not be exactly defined here.

5.2.3 Single point calibration

By carrying out a single point calibration, the factual
slope of the sensor can be established. The calibra-
tion medium can be water with known oxygen satu-
ration index (e.g. air-saturated water) or air with
known water-vapor saturation (e.g. water-vapor sat-
urated air).
After the sensor signal has stabilized, the complete
measuring system can then be calibrated to
the100 % value of the desired measurable variable,
e.g.100 % air, 20.95 % O
(77 °F) and normal pressure (see instruction manual
for the transmitter).

5.2.4 Dual point calibration

By carrying out a dual point calibration both slope
and zero point of the sensor can be established.
Important! In case of a dual point calibration,
always start by the zero point calibration before
calibrating the slope.
As a rule, the zero point should be adjusted to zero
manually, or it is automatically performed by the
transmitter (see instruction manual for the transmitter).
Zero-point calibration is recommended:
– when measuring less than 5 ppb in liquids
– when measuring less than 125 ppm (by vol.) in
gases
Attention! Incorrect zero point calibration is a
frequent source of measurement error. For correct
calibration, we recommend the use of carbon
dioxide gas with a level of purity of at least
99.9995 %.
After the sensor signal has stabilized (after 6 ... 12
hours), the sensor can be calibrated through the
relevant transmitter to the zero % value of the desired
measurable variable, e.g. 0 % air, 0.0 % O
0.0 ppm (see instruction manual for the transmitter).
© 02 / 09 Mettler-Toledo AG
Printed in Switzerland
Sensor 12 / 25 mm
2
, or 8.26 ppm at 25 °C
2
InPro 6850i, 6900i, 6950i
19
, or
2
52 206 349