Specifications
Flow Rates*
Dimensions
Weight
Sound Pressure Level
Power Consumption
O2 Concentration
Electrical Requirements
Operating Environment*
Altitude
Storage Environment
* Based on an atmospheric pressure range of 700 hPa to 1060 hPa at 70°F (21°C)
The expected service life of the equipment is a minimum of five years.
See Technical Manual (PN MN138-1) for Sound Power Level.
WARNING: USE OF THIS DEVICE AT AN ALTITUDE, TEMPERATURE, OR RELATIVE HUMIDITY
OUTSIDE OF THE SPECIFIED VALUES LISTED IN THIS MANUAL CAN ADVERSELY AFFECT THE
FLOWRATE AND THE OXYGEN CONCENTRATION, AND CONSEQUENTLY THE QUALITY OF THE
THERAPY.
WARNING: USE OF THIS DEVICE AT AN ALTITUDE ABOVE 10,000 FT (3048 M) OR OUTSIDE A TEM-
PERATURE OF 5° C TO 40° C (41° F TO 104° F ) OR A RELATIVE HUMIDITY ABOVE 95% IS EXPECTED
TO ADVERSELY AFFECT THE FLOWRATE AND THE PERCENTAGE OF OXYGEN AND CONSEQUENTLY THE
QUALITY OF THE THERAPY .
VisionAire
1 LPM – 5 LPM
±10% of indicated setting, or 200 mL, whichever is greater*
14.1 in. W x 11.5 in. D x 20.8 in. H
(35.8 cm W x 29.2 cm D x 52.8 cm H)
30 lbs (13.6 kg)
45 dBa (1 to 5 LPM)
290 Watts
90% (+ 5.5% / - 3%)
115 VAC / 60 Hz, 3.0A
230VAC /50 Hz, 1.5A
230VAC / 60 Hz, 1.5A
5° C to 40° C (41° F to 104° F ) at altitudes from up to 10,000 ft (3048
m) above sea level. 15 - 95% relative humidity (non-condensing)
-1250 to 10,000 ft (-381 to 3048 m) (tested to 700 – 1060 hPa)
-25° C - 70° C (-13°F - 158°F)
VisionAire Family
PN MN238-C4 A | User Manual
ENG - 13