Error code
E06
E07
E08
E10
E11
E12
Fault
Missing or Invalid Thermocouple
- Temperature probe not connected
or damaged
- Temperature probe removed
during operation
- Temperature probe still cooling
down from a previous heating job.
Power electronics failure
- Electronics disturbance
- Electronic damage
Mains supply voltage out of range
- Eventual voltage peak
- Power supply out of tolerance (10%)
Environment Temperature
out of range
- Operating range in between
0 - 40 °C (32 -104 °F)
No calibration data
- Electronics failure
Wrong voltage supply
- Voltage out of calibration range
Action to ix it
- Check Thermocouple connection
and wire damage
- If damaged, use Time Mode and
use an external thermometer
- Do not remove the probe during
operation.
- Allow some cooling time until
temperature probe reading
stabilization.
- Reset the heater
- Replace power print
(send to reseller)
- Reset the heater
- Check mains supply for correct
level
- Make sure the heater is within
temperature range
- Use it indoors and avoid sun
- Return for re-calibration
- Check the voltage version of the
heater
- Check the voltage of the power
supply