Herunterladen Inhalt Inhalt Diese Seite drucken

Pump Fixing; Technical Information; Emergency Stop - Varian TV 551 Navigator Bedienungshandbuch

Vorschau ausblenden Andere Handbücher für TV 551 Navigator:
Inhaltsverzeichnis

Werbung

PUMP FIXING

!
WARNING!
If a rotor failure occurs, the connection of the pump to the sys-
tem could be subjected to a significant torque. If the connection
is not sufficient to withstand that torque, the pump could detach
from the system or the motor housing could detach from the
pump envelope. In this case metal fragments could be pro-
jected from the pump or system, which could cause serious
injury or death and/or damage to surrounding equipment.
Fix the TV 551/701 Navigator in a stable position, mounting the
inlet flange of the turbopump to the system counter-flange, with
a connection capable of withstanding a torque of 8600 Nm
around its axis.
For example the ISO-K flange can be fixed using high strength
steel clamps (like Varian model IC63250DCMZ).
The following table shows, for each flange, the necessary
number of IC63250DCMZ clamps and the relevant fixing
torque.
FLANGE
N.
ISO 100
16
ISO 160
10
ISO 200
8
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Varian hard-
ware. See the appendix "Technical Information" for a detailed
description.
The TV 551/701 Navigator can be installed in any position.
NOTE
The TV 551/701 Navigator cannot be fixed by means of its
base.
!
The TV 551/701 Navigator belongs to the second installation
(or overvoltage) category as per directive EN 61010-1. Connect
the device to a mains line that satisfy the above category.
The TV 551/701 Navigator has Input/Output and serial com-
munication connectors that must be connected to external cir-
cuits in such a way that no electrical part is accessible.
Be sure that the insulation of the device connected to the TV
551/701 Navigator is adequate even in the case of single fault
as per directive EN 61010-1.
USE
This paragraph details the fundamental operating procedures.
Make all electrical an pneumatic connections before the use of
the system.
While heating the vacuum chamber, the temperature of the
inlet flange must not exceed 120 °C.
!
WARNING!
Never use the turbopump when the inlet flange is not con-
nected to the vacuum chamber.
Do not touch the turbopump or any of its accessories during
the heating process. The high temperatures may cause burns.
!
Avoid impacts, oscillations or harsh movements of the pump
when in operation. The bearings may become damaged.
Use air or inert gas free from dust or particles for venting the
pump. The pressure at the vent port must be less than 2 bar
(above atmospheric pressure).
For pumping aggressive gases, these pumps are fitted with a
special port to allow a steady flow of inert gas (like N
pump bearing protection (see the appendix "Technical Informa-
tion").
FIXING TORQUE
35 Nm
35 Nm
35 Nm
CAUTION
CAUTION
, Ar) for
2
!
When employing the pump for pumping toxic, flammable, or
radioactive gases, please follow the required procedures for
each gas disposal.
Do not use the pump in presence of explosive gases.
Switching on and Use of TV 551/701 Navigator
To switch on the TV 551/701 Navigator it is necessary to sup-
ply the mains. The integrated controller automatically recog-
nizes the interlock and start signals presence and start up the
pump.
The first pump start up is in "Soft Start" mode. When the start
up cycle is finished, the "Soft Start" mode automatically is dis-
abled, and the following start ups are without the "Soft Start"
mode. To re-enable the "Soft Start" mode it must be activated
by the suitable software command (see the paragraph "RS
232/485 COMMUNICATION DESCRIPTION" in the appendix

"Technical Information").

The green LED located on the TV 551/701 Navigator base front
panel indicates with its flashing frequency the system operating
conditions:
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in
ramp, or in braking, or in Stop, or in "Waiting for interlock"
status;
fast flashing (period of about 200 ms): error condition.
TV 551/701 Navigator Switching off
To switch off the TV 551/701 Navigator it is necessary to re-
move the mains. The integrated controller immediately stops
the pump.

Emergency Stop

To immediately stop the TV 551/701 Navigator in an emer-
gency condition it is necessary to remove the supply cable from
the mains plug.
MAINTENANCE
The TV 551/701 Navigator does not require any maintenance.
Any work performed on the system must be carried out by au-
thorized personnel.
!
Before carrying out any work on the system, disconnect it from
the mains, vent the pump by opening the appropriate valve,
wait until the rotor has stopped turning and wait until the sur-
face temperature of the pump falls below 50 °C.
In the case of breakdown, contact your local Varian service
center who can supply a reconditioned system to replace that
broken down.
Before returning the system to the constructor for repairs, or
replacement with a reconditioned unit, the "Health and Safety"
sheet attached to this instruction manual must be filled-in and
sent to the local sales office. A copy of the sheet must be in-
serted in the system package before shipping.
If a system is to be scrapped, it must be disposed of in accor-
dance with the specific national standards.
24
INSTRUCTION MANUAL
WARNING!
WARNING!
NOTE
87-900-934-01(D)

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Tv 701 navigator

Inhaltsverzeichnis