Herunterladen Inhalt Inhalt Diese Seite drucken

Specifications; Measurement.categories - Greenlee CJS-100 Bedienungsanleitung

Mit offener messgabel
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 43

Specifications

Display: 3-1/2-digit LCD (1999 maximum reading)
Sampling Rate: 1.5 per second
Overrange Indication: "OL" appears on the display
Jaw Opening: 16 mm (0.63")
Wireless Electric Field Detection (EF):
Voltage Range: 50 VAC to 1000 VAC
Frequency Range: 50 Hz to 500 Hz
Measurement Category: Category IV, 600 V; Category III, 1000 V per UL 61010-1
and UL 61010B-2-032
Temperature Coefficient: 0.2 x (specified accuracy) per °C
below 18 °C or above 28 °C
Operating Conditions:
At 0% ≤ 80% RH: 0 °C to 30 °C (32 °F to 86 °F)
At 0% ≤ 75% RH: 30 °C to 40 °C (86 °F to 104 °F)
At 0% ≤ 45% RH: 40 °C to 50 °C (104 °F to 122 °F)
Altitude: 2000 m (6500') maximum
Indoor use only
Storage Conditions: -20 °C to 60 °C (-4 °F to 140° F),
0% to 80% relative humidity with battery removed
Pollution Degree: 2
Battery: Two 1.5 V (AAA, A4M, or IEC LR03)

Measurement.Categories

These definitions were derived from the international safety standard for insula-
tion coordination as it applies to measurement, control, and laboratory equipment.
These measurement categories are explained in more detail by the International
Electrotechnical Commission; refer to either of their publications: IEC 61010-1 or
IEC 60664.
Measurement.Category.I
Signal level. Electronic and telecommunication equipment, or parts thereof. Some
examples include transient-protected electronic circuits inside photocopiers and
modems.
Measurement.Category.II
Local level. Appliances, portable equipment, and the circuits they are plugged into.
Some examples include light fixtures, televisions, and long branch circuits.
Measurement.Category.III
Distribution level. Permanently installed machines and the circuits they are hard-
wired to. Some examples include conveyor systems and the main circuit breaker
panels of a building's electrical system.
Measurement.Category.IV
Primary supply level. Overhead lines and other cable systems. Some examples
include cables, meters, transformers, and other exterior equipment owned by the
power utility.
CJS-100
1

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis