Herunterladen Diese Seite drucken

Agilent Technologies TwisTorr 84 FS X3502-64000 Bedienungshandbuch Seite 312

Werbung

17
Technical Information
Water Cooling Kit Connection
The assembled kit must be screwed into the suitable holes of the
pump body with a recommended closing torque of 5 Nm.
Cooling may be carried out either through an open circuit with
eventual discharge of the water, or using a closed circuit cooling
system.
The water temperature must be between +15 °C and +25 °C, with an
inlet pressure between 2 and 4 bar. This allows a flow higher than 65
l/h.
NOTE
The water electrical conductance must be ≤ 500 µs/cm. When the conductance
is higher, in closed water circuit, the use of up to 20 % of Ethyl-Glycole is
suggested.
312/336
TwisTorr 84 FS User Manual / 87-901-043-01(B)