VF Series Pallet Scales
5)
Connect wires
6)
Screw cover onto connection box and make scale ready for operation.
Note: The following tests must be performed after a load cell has been replaced:
Check side load and adjust as needed
Calibrate with maximum load
10.4
Replacing Levelling Feet
1)
Stand scale up
2)
Unhook retaining springs from the defective foot and remove it
3)
Lubricate new foot with O-ring and insert into the load cells
4)
Hook retaining springs into the new foot
11. SPARE PARTS
Analogue PCB
Levelling foot
Rubber for foot plate
Retaining spring
Connecting cable analog, 5m (M20x1,5)
Cable gland (M12)
Beam load cell 0,5t / C3 / for 600 kg platform
Beam load cell 1,0t / C3 / for 1500 kg platform
Seal (housing)
Shim for load cell 0,2 mm
Designation
Order Number
00205924
22015032
22003510
22003508
22009304
22008791
72192181
72192182
22015355
00204539
EN-8
Min. Order Quantity
1 piece
1 piece
1 piece
1 piece
1 piece
4 pieces
1 piece
1 piece
1 piece
4 pieces