Cleaning and care

The device reminds you to clean the measuring sensor af-
ter taking several temperature measurements. When this
occurs, the cleaning symbol
appears in the LC display
20
and acoustic signals are heard. In order to avoid con-
4
tamination with germs, the lens has to be rubbed after
every use with a soft cloth that has been moistened with
70% medical alcohol. The use of hygiene attachments is,
therefore, not necessary. After that, wait for at least
30 minutes before taking another measurement. Please
keep in mind that due to the evaporative cooling eff ect of
the alcohol, even the smallest amounts can falsify the result.
Clean the housing with a soft, slightly damp cloth. In the
case of more severe soiling a mild detergent can also be
added to the water.Keep the appliance in a dry environ-
ment and protect from direct sunlight. The appliance can
be stored at temperatures of –10.0 °C - 55.0 °C
(14.0 °F - 131.0 °F). Temperatures over 80.0 °C are to
be avoided! In order to measure with it, however, it is nec-
essary to let the appliance adjust to room temperature for
a suffi cient length of time - otherwise an error message
will be generated (see "Functions / Fault Analysis - Correc-
tion of Problems"). A room temperature of 5.0°C - 59.9°C
(41.0°F - 139.8°F) should also be maintained for the pur-
poses of measurement. If the temperature is outside this
range, an error message is shown (see "Functions / Fault
Analysis - Correction of Problems").
Remove the battery when not using the appliance for any
length of time.